Microprocessor Pressure Calibrator PC-01



  • Pressure measurements with 0,05% F.S. inaccuracy
  • Two pressure measurement scopes
  • Measurement of the standard electric signals with 0,02% F.S. inaccuracy
  • Simultaneous indications of:
  • pressure value
  • electric signal value for processing error characteristic of a verified converter (linear or radical)
  • Interface RS 232 (for measurement data transmission from a calibrator’s memory to a PC
  • Calibration certificate (upon a client’s request)
  • Possibility of choosing a language for displaying communicates and instructions (after agreeing it with a producer)
kalibrator

Designed for

Calibrator PC - 01 is a portable measurement device designed for calibration and verification of the following types of converters: pressure, pressure difference, manometers, pressure signalers, transmitters of the standard electric signals etc. The calibrator allows on performing the measurements both in the laboratories as well as in the object’s conditions.

Construction

The calibrator PC - 01 built-in a dustproof casing made out of the aluminum profiles. It consists of the following accessible (form the outside) elements: two pressure sensors terminals, the terminals RS 232, three sockets allotted for the electric signals, LCD display, a membrane keyboard, a power switch and a socket for battery charging. The keyboard contains twenty five key buttons that considering their functionality can be divided into four groups:

Technical data

PRESSURE MEASUREMENT
Scope of minimum measurement
0...100 kPa (minimum width of the measurement scope 30 kPa);
0...200 kPa (minimum width of the measurement scope 50 kPa);
0...700 kPa (minimum width of the measurement scope 200 kPa);
0...1700 kPa (minimum width of the measurement scope 500 kPa);
Scope of maximum measurement
0...3500 kPa (outside sensor)
Inaccuracy of readings
0...3500 kPa (outside sensor)
Stability of readings:
0,02% per year
Error of the temperature readings:
0,005 %/ C F.S.
Overload:
2 x measurement scope 2 x measurement scope 2 x measurement scope
Media measured:
gases and liquids indifferent to acid-resistant steel 316 ss
Electric signals measurements
Voltage measurement scopes:
0/4...20 mA dc;
Voltage measurement scopes:
0/1...5 V dc, 0/2...10 V dc
Inaccuracy of readings:
0,02% F.S.
Stability of readings:
0,01% per year
Error of the temperature readings:
0,002%/ C F.S.
Current indication scopes:
0...50 mA dc
Voltage indication scopes:
0...30 V dc
Calibrator’s memory capacity:
Storing a set of data and measurement results for 20 tested devices
POWER SUPPLY
battery
battery NiMH  
network a battery charger (in the set)
WORK CONDITIONS
Scope of temperatures 5..50C
Scope of humidity  5..95% (in the scope between 10...30C)
Storage temperature -40...60C
*) - Going beyond the work conditions (i.e. using the device in an open space) requires using additional protective means that secure the device’s electronic parts from water getting inside it.
- The device is produced basing on a license for the Industrial Research Institute for Automation and Measurements "PIAP" in Warsaw.

Courses

Effective work of the devices is undeniably influenced by a user’s understanding of the details concerning the device’s work principles and methods of selecting the correct parameters. The LAB-EL company organizes the training courses concerning the offered devices’ scope of service for the users, distributors and system maintenance staff.