THERMOMETER – HYGROMETER LB-701 with a reading panel
LB-705
|
|
DESCRIPTION
Thermometer - hygrometer LB-701 with a reading panel LB-705 is designed for air temperature and humidity measurement in locked places (warehouses, laboratories etc.). It is a stationary device, powered from 220V/50Hz network supply or from an outside source of 12V constant voltage.
The device is composed of two major parts: hytherograph LB-701 (made in a form of a measurement probe) and a reading panel LB-705. A cable 1 meter in length connects both parts. There is a possibility of extending the probe’s cable up to 14 m. Replaceable casings protect the measurement probe sensors (their type depends on what the device is used for). The device in a standard version is equipped with s sensor casing (anti pollution filter) type F1.
The measurement probe LB-701 includes: a thin layer temperature
sensor Pt-1000, thin layer capacitive relative humidity sensor,
electronic processing systems and a system of programmed non volatile
memory, where all individual probe characteristic are stored as well as
identification and technical data.
The reading panel receives signals that had been already processed,
which include coded information concerning temperature and relative
humidity values. It allows on changeable use of different reading
panels with the same hytherograph LB-701 without any influence on the
measurement results.
The microprocessor from the device's reading panel, basing on the
information from the probe (processed measurement signals and read
calibration characteristics) computes the following values: relative
humidity, temperature, dew point temperature and volume water vapor
content in the air. The temperature and relative humidity values are
presented on the alphanumeric LCD screen at the same time (the size of
numeric characters 25 mm) and can be sent out through the interface.
The lack of switches increases device's reliability at work and makes
its use a lot easier.
The device is equipped with two serial interfaces: RS232C, which allows for it to be attached to any computer system and with a LAB-EL digital current loop for data transmission to long distances. The interfaces allow on remote and automatic measurement data collection and entering the calibration data during the original and secondary calibration.
Panel LB-705 has a programmed local time clock, which counts the
time from the moment of turning on the power supply, or synchronizes it
with the time real equipment clock. The panel has a non-volatile memory
of measurement results (capable of storing approximately 640
measurement points) what along with a real time clock allows on using
it as a temperature and humidity recorder. An interval between the
entries can be set from 1 up to 90 minutes (with a 1 minute stop) or
from 100 to 1580 minutes (with a 10 minute stop). The device is quipped
with a PC computer program used for setting the registry parameters and
for reading the memory.
The device does not have any mechanic calibration elements (i.e. potentiometers). Sending the calibration data obtained in model climate conditions trough a digital interface to device’s non-volatile memory performs the device’s calibration. It guarantees comfort and calibration accuracy as well as stability of device's meteorological parameters in time.
Thermometer - hygrometer LB-701 obtained a certificate type RP T 95
76 issued by the President of the Central Measurement Office in Warsaw.
Thermometer- hygrometer LB-701 received a positive expertise from The Central Army Meteorology Office in Zielonka (no. 71 dated 05.05.1995), performed basing on the method of legalizing hygrometers and hygrographs.
Each thermometer-hygrometer, upon a special order, can have an individual calibration certificate from the LAB-EL Calibration Laboratory.
The device is produced also in a version for
temperature
measurement (thermometer LB-701T with the panel LB-705).
Thermometer LB-701T has a measurement converter (in a cylinder case)
and an outside temperature sensor Pt1000 (or Pt100) in a standard or
special case, adjusted to the individual user requirements (i.e. in a
form of a rod from the acid resisting steel, with a special cable).
Special version of the sensor casing allows
on different values
measurement in an extended scope of temperatures (-200...+550°C).
Thermometer LB-701T can be also purchased separately, as additional
equipment for the set LB-701/LB-705. However, in such case it is
required that the internal software for the older versions of panel
LB-705 is updated. Thermometer LB-701T cable length amounts to: between
an outside temperature sensor and a converter - 1..40 m, between a
converter and a reading panel - 1..15 m (the cables length is defined
by the customer).
Methods used for relative humidity and air temperature measurements using hytherographs are introduced in a description below.
TECHNICAL DATA
| TEMPERATURE
MEASUREMENT |
||
|---|---|---|
| Measurement inaccuracy |
+/- 0,1°C |
+/- 1 last digit |
| Scope of measurement | -40.. +85°C -200.. +550°C |
LB-701 LB-701T - sensor in a special casing |
| HUMIDITY
MEASUREMENT |
||
|---|---|---|
| Measurement inaccuracy |
+/- 2,0% +/- 4,0% |
in the scope of 10.. 90%,
+/- 1 last digit outside of the scope 10.. 90% |
| Scope of measurement LB-701 |
10.. 95% 10.. 60% 10.. 40% |
for temperatures up to +40°C for temperatures up to +70°C for temperatures up to +80°C |
| Scope of measurement LB-701H | 0.. 100% 0.. 70% 0.. 40% |
for temperatures up to +60°C for temperatures up to +70°C for temperatures up to +80°C |
| MEASUREMENT READINGS | ||
|---|---|---|
| Parameter |
Scope of readings |
Resolution |
| Air temperature (LB-701) |
-40,0.. +85,0°C | 0,1°C |
| Air temperature (LB-701T) |
-199,9.. +550°C |
0,1°C |
| Relative humidity |
0,0.. 100,0% |
0,1% |
| MEASUREMENT
READINGS (additional
using a series interface) |
||
|---|---|---|
| Parameter |
Scope of readings |
Resolution |
| Dew point temperature |
-40.0..+100.0°C |
0.1°C |
| Volume water vapor content in
the air |
0..99999ppm |
1 ppm |
| Time |
months, days, hrs., min |
1 min |
| SCOPE OF WORK TEMPERATURES | ||
|---|---|---|
| Device |
Scope |
Comments |
| Probe LB-701M |
-40,0.. +85,0°C |
|
| Thermometer sensor LB-701T |
-60..+90°C -200..+550°C |
for a standard version for a special version |
| Thermometer converter LB-701T |
-40,0.. +85,0°C |
|
| Panel LB-705 |
0..+50°C |
|
| RECOMMENDED CONSTANT WORK CONDITIONS *) | |
|---|---|
| Scope of temperatures |
10..40°C |
| Scope of humidity | 20..80% |
| Level of environment corrosion
aggressiveness accord. PN-71/H-04651 |
B |
| Degree of casing protection |
IP 60 (limited additionally by
the class of filter protection) |
| POWER SUPPLY | ||
|---|---|---|
| Network | 230V AC / 50Hz / 10VA |
|
| Low voltage | feeder 12V DC / 0.4A | |
INTERFACE I
Series RS232C, 3 lines (RxD, TxD, GND), transmission parameters: 9600
bps, 8 bits of information, without the evenness control, 1 stop bit.INTERFACE FUNCTIONS I
Creates a possibility of reading the measurement results and storing of the device's calibration in its non-volatile memory.INTERFACE II
Series so called digital current loop, 1 line TXD (active TXD line
stage complies with 25 mA current, in the rest stage 15 mA), time
transmission parameters in accordance with RS232C: 300 bps, 7 bits of
information, without evenness control, 1 stop bit.INTERFACE FUNCTIONS II
Possibility of sending automatically the measurement results to long distances in big measurement systems (i.e. to multi sensor climate supervision systems - see LB-731). Galvanic isolation output.OVERALL DIMENSIONS
- Reading panel LB-705: in a panel case (to be hanged on the wall) with dimensions: 140 x 190 x 116 mm.
- Thermometer - hygrometer LB-701 (measurement probe): attached with 1 meter cable, with dimensions: diameter 23 mm, length (with a stay wire) 190 mm (for a probe’s metal casing: diameter 18 mm, length 150 mm).
- Thermometer LB-701T in a standard version: converter in a cylinder case with diameter 18 mm and length 160 mm, sensor in a stainless steel case with diameter 5 mm and length 100 mm, ended with a blade (see the photograph above).
EQUIPMENT
- cable for the interface RS232C,
- program "User's Interface" for a PC computer.
EXTENDED VERSIONS
The following effects are available upon special order:- LB-701H - with extended compared to a standard, range of constant humidity measurement to 0...100%,
- LB-701T - thermometer (converter with a sensor) for air temperature or any value measurement (made according to specific client requests),
- LB-701Fn - sensor casing version (according to a description), the version is given if it is different than F1.