THERMOMETER- HYGROMETER LB-710L (LB-710/ML/HL/AL)
|
![]() |
Description
Hytherograph LB-710L is designed for air temperature and relative humidity measurement in closed places (museums, warehouses, archives, laboratories). It is a stationary device, of small size, powered from an outside power supply.
For measurement the device uses data from the temperature sensor Pt-1000 as well as from the capacitive relative humidity sensor. Basing on the knowledge of nonlinear sensor characteristics and digital calibration data stored during the calibration process in a device's memory - the microprocessor (built-in the device) counts the actual relative humidity and temperature measurement results. Those results are sent (in a digital form) from the device to the supervisory system of data collection.
The device is equipped with a digital current interface and its time course is analogous to the standard RS232C. Such interface beside data transmission is also used for powering the hytherograph - using two-cable connection for this purpose (i.e.: telephone type), but the polarity of attaching the interface's cables is optional.
The device can be attached to any supervisory system using the interface i.e.: climate regulator LB-722, concentrator LB-731, panel LB-755 or (using a converter LB-371 or LB-486) to an interface RS232C of any computer. It allows on remote control and automatic measurement data collection.
The device does not have any mechanical calibration elements (i.e.: potentiometers). The calibration of the device is performed by sending calibration data obtained in the model climate conditions through a digital interface to the non-volatile memory of device. It guarantees comfort and high calibration accuracy as well as stability of the device's meteorological parameters in time.
Each hytherograph (upon a request)
can have an individual calibration
certificate issued by the LAB-EL Calibration Laboratory.
Thermometer LB-710 received the certificate
of type RP T 95 77 issued
by the President of Central Measurement Office in Warsaw.
Technical data
| TEMPERATURE MEASUREMENT | ||
|---|---|---|
| Measurement inaccuracy |
+/- 0.2°C | +/- 1 last digit |
| Scope of measurement | -40..+85°C | version LB-710L |
| HUMIDITY MEASUREMENT | ||
|---|---|---|
| Measurement inaccuracy |
+/- 2.0% | in the scope of 10...90%, +/- 1 last digit |
| +/- 4.0% | outside of the scope 10...90% | |
| Scope of measurement LB-710L | 10..95% | for temperatures up to +40°C |
| 10..60% | for temperatures up to +70°C | |
| Scope of measurement LB-710HL | 0..100% | for temperatures up to +60°C |
| 0..70% | for temperatures up to +70°C | |
| MEASUREMENT READINGS |
||
|---|---|---|
| Parameter |
Scope of readings | Resolution |
| Air temperature |
-40.0..+85.0°C | 0.1°C |
| Relative humidity | 0.0..100.0% | 0.1% |
| RECOMMENDED CONSTANT WORK CONDITIONS *) | |
|---|---|
| Scope of temperatures |
0..40°C |
| Scope of humidity | 20..80% |
| Level of environment corrosion aggressiveness accord. PN-71/H-04651 | B |
| Degree of casing protection |
IP 20 |
| POWER SUPPLY |
||
|---|---|---|
| Scope of voltage | 8..24V DC | |
| Energy consumption | 3mA | average |
| 25mA | maximum |
|
INTERFACE
Transmission parameters: time in accordance with RS232C, 300 bps, 7 information bits, without the evenness control, 1 stop bit working in a standard of a digital current loop.OVERALL DIMENSIONS
Casing type L: approx. 70 x 70 x 22 mm.EXTENDED VERSIONS
The following versions are available upon a special order:- LB-710HL - with extended compared to standard scope of constant humidity measurement to 0...100%.
- LB-710AL - with increased temperature measurement accuracy.