Time constants of the temperature sensors


  • Temperature measurement result depending on: 
    • diameter and width of a sensor’s wall
    • temperature jump
    • speed of mixing
Stałe czasowe czujników

Temperature measurement result obtained by using an electronic thermometer with an outside temperature sensor, in a casing made out of a stainless steel pipe, becomes stabilized with certain speed resulting from:

It means, that after placing a sensor in a measured medium (air, liquid, refined material etc.), which temperature is different from a sensor’s temperature by a value “delta u” (on a graph’s vertical axis), the sensor’s temperature will gradually (unfortunately not immediately) get close to the temperature of a measured medium. A typical course of changes as a result of temperature measurement was introduced on a graph in a time function (on a horizontal axis). Characteristic times for the chosen sensor sizes and for two different media (water and air) were provided in the table below. For different media - time constants usually are going to stay between data for those two media mentioned in the table.

The fact that is worth mentioning is that: 

Stałe czasowe czujników

THE SENSORS WITH THE SHIELDS MADE OUT OF THE ACID RESISTANT STEEL PIPES 1H18N9T DIMENSION
(pipe diameter x wall’s thickness)
CHARACTERISTIC TIMES (in seconds)
water, flow 0,4  m/s air, flow 1,0 m/s
T 0,5
T 0,9
T 0,5
T 0,9
3 x 0,25
7
19
35
95
4 x 0,35
12
32
45
125
5 x 0,35
17
45
70
190
6 x 0,35
22
61
80
235
8 x 0,60
27
72
105
310
10 x 1,60
35
96
140
375
12 x 2
51
138
165
420
Elektronika Laboratoryjna Producent: LAB-EL Elektronika Laboratoryjna Sp. J.
ul. Bodycha 68b, 02-495 Warszawa 
tel: (022) 667 71 18, 0601 209656, fax: (022) 867 53 32 
poczta elektroniczna: info@label.com.pl
witryna internetowa: http://www.label.com.pl
internetowa stacja meteorologiczna: http://www.meteo.waw.pl
stacja meteorologiczna w komórce: wap.meteo.waw.pl
Ostatnia aktualizacja: 23.06.2007
iso9001 pomiary